Open original FDA record ↗Recall record
- FDA identifier
- Z-0045-2012
- Company in FDA record
- Interventional Spine Inc
- FDA classification date
- Oct 13, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
This recall has been initiated due to drivers that may contain a burr in the distal tip preventing full seating of the driver onto the head of the facet screw, resulting in difficulty screwing the facet screw into the bone. Use of this product may result in a delay of the surgical procedure
Source facts
- Action
- Interventional Spine, Inc. sent an "Urgent Recall" letter dated March 15, 2011 to all affected customers. The letter included; affected product, reason for recall and recommendations. Customers were asked to immediately examine inventory, quarantine affected product and to notify International Spine, Inc if any product remain. For information on this recall contact the firm at (949) 472-0006.
- Initiated
- 2011-02-03
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-02-05
- Distribution
- Word wide Distribution: (USA) Nationwide including the states of CO, CT, FL, KY, LA, MA, NC, NY, OK, OR, TN, TX, VA, and WI; and the countries of Italy, Spain, Turkey, and the U.K.
- Recall number
- Z-0045-2012
- Recall status
- Terminated
- CDRH record ID
- 99401
- FDA classified
- 2011-10-13
- Recall event number
- 58461