The ROI-A Anterior Delivery Device T-Handle, part number SI-ROIA-0023 has experienced binding of the rotation of the t-handle. There have been no reports of patient injury
Source facts
Action
The firm, LDR, contacted affected consignees via phone and then sent a follow up email dated 9/8/16. The phone script and email described the product, problem and actions to be taken. The consignees were instructed to promptly return the instrument using the RMA#1180. The firm will perform a rework that prevents the binding to occur.
If you have any question, contact the Quality Engineering Manager by email: ron.musselman@zimmerbiomet.com, phone: 512-344-3436 or Field Inventory Manager - US Supply Chain or phone: 512-344-3436.
Initiated
2016-08-29
Root cause
Device Design
Terminated
2017-01-13
Distribution
US Distribution to states of: GA, OR, FL, KY, NC, AR and PA.
Recall number
Z-0045-2017
Recall status
Terminated
CDRH record ID
149274
FDA classified
2016-10-05
Recall event number
75068