Skip to content

Public FDA record

Z-0045-2018: Sysmex Fluorocell(R) PLT Fluorocell PLT is used to stain platelets in a diluted blood sample for the cou

Recall record Z-0045-2018 for Sysmex America, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0045-2018
Company in FDA record
Sysmex America, Inc.
FDA classification date
Oct 16, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is potential that fluorescent platelet (PLT-F) and immature platelet fraction (IPF%, #) results obtained using the lots listed above may exhibit lower than actual PLT-F and IPF values. The issue may be recognized by a large discrepancy between the impedance platelet (PLT-I) result and the PLT-F result. In most cases, the falsely decreased PLT-F results display a PLT Abn Scattergram message with an asterisk beside the result or an Action Message: Difference between PLT and PLT- F. Check the results. These are indicators of unreliable data. In rare instances, platelet flags may be absent. The PLT-I results are unaffected.

Source facts

Action
Sysmex sent a Customer Letter dated August 2017 to all affected customers. The customer was directed to examine their inventory for the affected lot numbers and report back to Sysmex. .Customers with questions should call 1-888-879-7639, option 2 (if within the U.S.) or 1-888-679-7639 (for customers in Canada)
Initiated
2017-09-08
Root cause
Nonconforming Material/Component
Distribution
Worldwide Distribution - US (nationwide including Puerto Rico) Internationally to New Zealand, Brazil, India, Philippines, Nicaragua, Argentina, Australia, Chile, Viet Nam, Colombia, Pakistan, Malaysia, Dubai, Germany, Hong Kong, Bangladesh, Uruguay Thailand, China, Saudi Arabia, Taiwan, Honduras, Ecuador, Kuwait, South Korea, and Costa Rica
Recall number
Z-0045-2018
Recall status
Completed
CDRH record ID
158748
FDA classified
2017-10-16
Recall event number
78134

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.