Z-0045-2025: Check-Flo Performer Introducer Set, Reference Part Number RCF-5.0-35-45-J, Order Number G27242. Performer

Recall record Z-0045-2025 for Cook Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0045-2025
Company in FDA record
Cook Incorporated
FDA classification date
Oct 9, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Cook Medical is conducting removal of a lot of Check-Flo Performer Introducer Set due to complaints that the white caps that connect the shaft to the hub of these products may disconnect.

Source facts

Action
An URGENT: MEDICAL DEVICE RECALL notification letter dated 9/13/24 was sent to customers. Actions to be Taken by the Customer 1. Examine inventory immediately to determine if you have affected product s from the lot listed above. If so, quarantine any affected products that remain unused. Immediately cease all further distribution and use o f the affected products. 2. Return the affected product s to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter regardless of whether you have affected product(s) on hand and email the completed form to ausrecalls@cookmedical.com. 4. Please share this notice with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected products have been transferred. 5. Please immediately report adverse events to Cook Medical Customer Relations by phone at 1800 777 222 or by email at cau.complaints@cookmedical.com .
Initiated
2024-09-16
Root cause
Process control
Distribution
International distribution to the country of Australia.
Recall number
Z-0045-2025
Recall status
Open, Classified
CDRH record ID
210141
FDA classified
2024-10-09
Recall event number
95405

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗