Z-0046-2012: Reusable BONE-LOK PLS System; Catalog #'s: 9039-01 and 9039-02. Interventional Spine, Inc. A sterile kit

Recall record Z-0046-2012 for Interventional Spine Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0046-2012
Company in FDA record
Interventional Spine Inc
FDA classification date
Oct 13, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

This recall has been initiated due to drivers that may contain a burr in the distal tip preventing full seating of the driver onto the head of the facet screw, resulting in difficulty screwing the facet screw into the bone. Use of this product may result in a delay of the surgical procedure

Source facts

Action
Interventional Spine, Inc. sent an "Urgent Recall" letter dated March 15, 2011 to all affected customers. The letter included; affected product, reason for recall and recommendations. Customers were asked to immediately examine inventory, quarantine affected product and to notify International Spine, Inc if any product remain. For information on this recall contact the firm at (949) 472-0006.
Initiated
2011-02-03
Root cause
Nonconforming Material/Component
Terminated
2013-02-05
Distribution
Word wide Distribution: (USA) Nationwide including the states of CO, CT, FL, KY, LA, MA, NC, NY, OK, OR, TN, TX, VA, and WI; and the countries of Italy, Spain, Turkey, and the U.K.
Recall number
Z-0046-2012
Recall status
Terminated
CDRH record ID
99402
FDA classified
2011-10-13
Recall event number
58461

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗