Open original FDA record ↗Recall record
- FDA identifier
- Z-0046-2014
- Company in FDA record
- Abbott Medical Optics, Inc.
- FDA classification date
- Oct 22, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
This corrective action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure of a thyristor component on the AC Power Distribution Printed Circuit Board (PCB).
Source facts
- Action
- The firm, Abbott Medical Optics, Inc., (AMO), sent an "ADVISORY NOTICE and a COVER LETTER FOR ADVISORY NOTICE" dated September 16, 2013 to all affected customers. The letters identified the product, problem, and actions to be taken by the customers. The customers were instructed to complete and return the Star Notification Acknowledgement form via fax to 1-714-247-4530 within 3 business days or mail to Abbott Medical Optics, 510 Cottonwood Drive, MS 347118, Milpitas, CA 95035, and share this advisory notice with your staff. If you have any questions related to operation of STAR Excimer Laser System or documentation requirements associated with this letter, please call AMO Technical Support at 1-800-511-0911 and you will be directed to an AMO Representative.
- Initiated
- 2013-09-16
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-11-12
- Distribution
- Worldwide distribution: US (nationwide) including Puerto Rico and countries of: Asia Pacific, Canada, Europe, Africa, Middle East, Japan, and Latin America.
- Recall number
- Z-0046-2014
- Recall status
- Terminated
- CDRH record ID
- 121965
- FDA classified
- 2013-10-22
- Recall event number
- 66304