Z-0046-2017: Toshiba America Medical Systems, Inc Aquilion Lightning TSX-035A CT Scanner.

Recall record Z-0046-2017 for Toshiba American Medical Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0046-2017
Company in FDA record
Toshiba American Medical Systems Inc
FDA classification date
Oct 6, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It has been found that if the scan range extends beyond the maximum Field of View (FOV), a reconstruction operation error may occur. As a result, scanning may be interrupted and rebooting of the system may be necessary. Please note that this issue has not occurred in the U.S.

Source facts

Action
Toshiba America Medical System, Inc. planned action to bring the defect into Compliance: 1. A notification will be sent to customers with affected systems in compliance with 21 CFR 1003.21 upon approval by the Secretary. The notification letter includes a statement by the correction will be made at no cost to the customer. 2. To correct this issue, modified software will be installed on the systems to prevent this occurrence from happening. The consignee will be contacted by their Toshiba service representative for an appointment when the revised software becomes available. 3. To avoid this problem in scan planning, the consignee will be advised to do the following until the modified software is installed: a. Confirm that the scan range does not extend beyond the maximum FOV. b. If the ROI for the scan range (indicated by the yellow solid lines) is set as the same size as the maximum FOV (indicated by green dashed lines), the ROI position must be changed. Instructions for changing ROI will be provided as attachments to the notification letter. 4. The modification to the software will be completed within 90 days from the date that the letters are submitted to Toshibas customers. 5. Progress letters will be provided to the Secretary, including the number of electronic products repaired. For further questions, please call (714) 730-5000.
Initiated
2016-04-25
Root cause
Software design
Terminated
2017-07-13
Distribution
US Distribution to the states of: PA, IA, MA, SD, MI, LA, KY, NC, NY, AR, IN, WI, OH, TY, FL, WV, WA, CT, MS, AL, ND, CO, HI, NO, MN, ID, OH, GA, MT, LA, NM, VT, TN, CA, NH, NJ, NE, SC and KS.
Recall number
Z-0046-2017
Recall status
Terminated
CDRH record ID
148643
FDA classified
2016-10-06
Recall event number
74871

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗