Open original FDA record ↗Recall record
- FDA identifier
- Z-0047-2007
- Company in FDA record
- Respironics California Inc
- FDA classification date
- Oct 19, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
This action is being initiated due to power supply snubber board failures on certain Esprit Ventilators in countries that utilize operating voltages of 200-240 Volts AC and will involve the replacement of the Esprit Ventilator power supply snubber board. Returned power supplies have shown evidence of overheating and in some cases in which a fan also failed, have shown signs of fire damage.
Source facts
- Action
- The field action will be conducted through consignees (distributors) in the affected countries. The distributors were notified of the field action by Federal Express au on August 8th, 2006. The distributors have been informed that Respironics will provide replacement circuit boards at no cost and will be instructed to upgrade the ventilator power supplies in their region. Instructions have been provided to the distributors and users on how to evaluate the ventilator to be able to continue to use it pending the completion of the upgrades. The distributors will document the replacement of the affected component by signing and dating the activity on the Ventilator Field Action Log (reference sample on Appendix IV). The serial number of the power supply destroyed will also be recorded on the log.
- Initiated
- 2006-08-08
- Root cause
- Other
- Terminated
- 2010-01-08
- Distribution
- Worldwide, including Algeria, Argentina, Australia, Bangladesh, Brazil, Cambodia, Chile, Egypt, France, Hong Kong, India, Italy, Jordan, Lebanon, Malaysia, Pakistan, Peru, Philippines, Russia, S Africa, S Korea, Spain, Thailand, Turkey, United Kingdon, & Vietnam
- Recall number
- Z-0047-2007
- Recall status
- Terminated
- CDRH record ID
- 48111
- FDA classified
- 2006-10-19
- Recall event number
- 36161