Open original FDA record ↗Recall record
- FDA identifier
- Z-0047-2015
- Company in FDA record
- Covidien LLC
- FDA classification date
- Oct 15, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Covidien Medi-Trace Cadence and Kendall Defibrillation Electrodes Not Compatible for use with Philips model FR3 or FRx AEDs
Source facts
- Action
- Covidien notified customers by letter via Federal Express on 9/18/14. The letter requests that they immediately verify that they are not stocking Covidien electrodes for use with Philips FR3 or FRx AED units. A customer reply acknowledgement form is provided to allow Covidien to track the effectiveness of the notification. Covidien is adding a label to all levels of packaging that alerts the user that the product is not compatible with the FR3 and FRx AED units. Distribution partners notified to address product in the distribution chain that is available to be relabeled. For questions customers should call (800) 962-9888, option 8, then extension 2500. For questions regarding this recall call 508-261-8000.
- Initiated
- 2014-09-18
- Root cause
- Device Design
- Terminated
- 2017-05-24
- Distribution
- Worldwide Distribtion-USA (nationwide) including Puerto Rico and the countries of Canada Greece Israel Belgium Finland France Germany Great Britain Ireland Italy Norway Poland Portugal Spain Sweden Switzerland Brazil Chile Philippines Taiwan Australia New Zealand.
- Recall number
- Z-0047-2015
- Recall status
- Terminated
- CDRH record ID
- 130112
- FDA classified
- 2014-10-15
- Recall event number
- 69257