Skip to content

Public FDA record

Z-0048-2018: GORE TAG, THORACIC ENDOPROSTHESIS, CONFORMABLE. The GORE TAG Thoracic Endoprosthesis is intended for en

Recall record Z-0048-2018 for W L Gore & Associates, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0048-2018
Company in FDA record
W L Gore & Associates, Inc.
FDA classification date
Oct 25, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Incomplete and/or partial deployment of the CTAG Device during the endovascular procedure.

Source facts

Action
The firm contact their consignee with a safety notice by letter on 09/25/2017. Gore is updating its Instructions for Use (IFU) to include the following new warnings and precautions: " If abnormal or inconsistent deployment line resistance is felt during deployment initiation, STOP deployment action immediately. If device remains constrained, remove device through the introducer sheath. If resistance is felt during removal through the sheath, stop and withdraw device and introducer sheath together. " If the device is in a partially deployed state and remains attached to the catheter, physicians should strongly consider conversion to immediate open surgical repair to avoid additional procedure time and potential harm from additional endovascular maneuvers.
Initiated
2017-09-25
Root cause
Unknown/Undetermined by firm
Terminated
2019-08-27
Distribution
Worldwide.
Recall number
Z-0048-2018
Recall status
Terminated
CDRH record ID
158746
FDA classified
2017-10-25
Recall event number
78132

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗