The x-ray tube head assembly of the Phantom Portable X-Ray System may detach from the boom arm assembly. The tube head could fall and contact the patient or operator.
Source facts
Action
Del Medical sent Field Correction Recall letters dated 10/3/06 to the direct accounts, informing them of the potential for the shoulder bolt in the yoke assembly to come loose. If the shoulder bolt comes loose, the tube head could potentially fall and contact the patient or operator. The accounts were requested to locate the affected systems and take them out of service until they are inspected and, if needed, repaired.
Initiated
2006-10-03
Root cause
Other
Terminated
2008-01-24
Distribution
Nationwide and Thailand
Recall number
Z-0049-2007
Recall status
Terminated
CDRH record ID
48715
FDA classified
2006-10-19
Recall event number
36478