Skip to content

Public FDA record

Z-0049-2018: Ablatherm(R) Integrated Imaging, High Intensity Ultrasound System For Prostate Tissue Ablation Transrec

Recall record Z-0049-2018 for Edap Technomed Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0049-2018
Company in FDA record
Edap Technomed Inc.
FDA record creation date
Oct 26, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The US FDA has requested the optional energy treatment settings, "medium" and "low" (Software Protocols 02-Medium and 03-Low) be removed from all Ablatherm Integrated Imagining devices in the U.S. until supporting clinical data can be submitted and evaluated by FDA.

Source facts

Action
The firm, EDAP TMS, company contacted all affected customers by phone during the week of August 14, 2017 and followed up with an " URGENT FIELD SAFETY NOTICE" letter dated 8/22/2017 delivered by email on 08/29/2016. The letter described the product, problem and actions to be taken. The customers were instructed as follows: do not use the "medium" or "low" energy treatment power settings in the meantime.The devices will be serviced at the consignee to remove Protocols "02-medium" and "03-low". If you have any questions regarding this notification, or experience any issues with your device, please contact EDAP TMS Customer Care Center at (512) 852-9685 or via email at service@edap-usa.com at any time, 24 hours a day, 7 days a week.
Initiated
2017-08-14
Root cause
Other
Terminated
2020-08-04
Distribution
US Distribution to states of: NY, FL CA, NC, TX and NJ.
Recall number
Z-0049-2018
Recall status
Terminated
CDRH record ID
158643
FDA record created
2017-10-26
Recall event number
78092

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗