Z-0049-2020: CareFusion MaxPlus Clear Needleless Connector. Intended for Positive Displacement Connector for needleles

Recall record Z-0049-2020 for Becton Dickinson & Company. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0049-2020
Company in FDA record
Becton Dickinson & Company
FDA record creation date
Oct 7, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

After disconnection of a luer, the valve of the connector remains recessed; not fully closing, either temporarily or permanently. The risks of blood loss and air embolism exist if the valve remains permanently open and goes undetected, especially in the neonate population.

Source facts

Action
BD MDS (Medication Delivery Solutions) initiated communication to US on May 2, 2019 via Fed'X. Letter states product defect, health risk and action to take:1. Immediately review for the specific Catalogue (Ref) and lot numbers . Destroy all product subject to the recall following your institutions process for destruction. 2. Share this recall notification with all users of the product within your facility to ensure that they are also aware of this recall. 3. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification and process your product replacement. Regional Customer Quality 888-237-2762 OPT 3, OPT 2 Monday Friday between the hours of 8:00am and 5:00pm (CST)
Initiated
2019-05-02
Root cause
Under Investigation by firm
Terminated
2021-11-30
Distribution
Nationwide Foreign: Canada
Recall number
Z-0049-2020
Recall status
Terminated
CDRH record ID
173255
FDA record created
2019-10-07
Recall event number
82970

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗