Skip to content

Public FDA record

Z-0049-2025: KINOVA Jaco assistive robot. Model Numbers: PJ 0000 0001, PJ 0000 0012, PJ 0090 0001, PJ 0090 0006, KR MJ

Recall record Z-0049-2025 for Kinova Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0049-2025
Company in FDA record
Kinova Inc
FDA classification date
Oct 11, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Potential installation issue that could result in a fire hazard in the event of an interaction between a damaged Jaco arm and a wheelchair exhibiting electrical leakage.

Source facts

Action
On 07/22/2024, Kinova notified consignees via UGENT: MEDICAL DEVICE CORRECTION letter. Consignees were instructed contact Kinova immediately for an assessment, ensure that if the wheelchair is in contact with a damaged part of the Jaco arm in any position at any time, to unplug the Jaco arm from the power source immediately and contact Kinova, and read the updated user guide on Kinova's website. Consignees were also requested to complete and return the acknowledgement and response form. Distributors were instructed to communicate the recall to all Jaco users, confirm receipt and understanding of the recall communication, maintain a detailed list of users, provide the user's contact details to Kinova, and perform an assessment for each user that response to the notification. Distributors were also instructed to complete and return the acknowledgement and response form.
Initiated
2024-08-22
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution in the states of California, Missouri, North Carolina, Ohio and the countries of Australia, Canada, China, Czechia, France, Germany, Netherlands, Norway, Portugal, Switzerland, Great Britain, Northern Ireland.
Recall number
Z-0049-2025
Recall status
Open, Classified
CDRH record ID
209236
FDA classified
2024-10-11
Recall event number
95065

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗