Skip to content

Public FDA record

Z-0050-2014: Solea 1.0 Laser Surgical instrument for use in general and plastic surgery and dermatology.

Recall record Z-0050-2014 for Convergent Dental. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0050-2014
Company in FDA record
Convergent Dental
FDA classification date
Oct 22, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Specific error condition could lead to unintended laser emission during laser activation.

Source facts

Action
The following Notice of Correction letter was sent out to all customers via email on 9/12/13, and followed up with a direct phone call to confirm receipt and review the contents of the letter: Dear Dr. [Customer name], This email is to inform you of a pending notice of correction regarding your Solea system, serial #[Customer serial number]. We have identified a specific error condition that could lead to unintended laser emission during laser activation. The software defect that caused the error condition has been identified and tested. Please do not use your Solea system until further notice. Resolution of this issue will require the installation of an updated release of the Solea software. A representative from Convergent Dental will be in touch with you to schedule the upgrade. Convergent Dental is committed to product quality and patient safety. If you have questions about this issue, please contact us. Software fix and testing: New software version (Release 2.0.4.0) was created by the software development group. Convergent Dental has implemented two software changes to eliminate this problem. The first is a defect fix to correct the handling of fatal errors, specifically laser communication timeout errors, so that these errors are not able to be cleared. This will prevent the user from being able to re-enter the workflow after a fatal error. The second is a defect fix to not set the laser state to disconnected when a laser timeout error condition occurs. This will ensure that the command to stop laser emission is always sent to the laser control module anytime the laser cutting state is exited due to error condition or the normal workflow event of stepping off the foot pedal. The software development team ran full Unit and Integration Tests on the new software build. The software verification activities included execution of the full verification test plan (Doc #790-00001), which has been revised to include specific defect
Initiated
2013-09-12
Root cause
Software design
Terminated
2013-11-15
Distribution
US distribution including states of: MA, CT, NY.
Recall number
Z-0050-2014
Recall status
Terminated
CDRH record ID
122154
FDA classified
2013-10-22
Recall event number
66361

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗