Z-0050-2025: StealthStation S8 App versions 1.3.0 and 1.3.2, (Model # 9735762) The StealthStation System, with Stealt

Recall record Z-0050-2025 for Medtronic Navigation, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0050-2025
Company in FDA record
Medtronic Navigation, Inc.
FDA classification date
Oct 10, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to two software anomalies that can occur within a spine procedure. Two issues are: 1. Navigation Orientation - when lateral orientation radiographic images are taken, information may appear flipped when utilizing the system (180 degree flipped). 2. Pop-up Message - when using CT+Fluouroscopy images with a slice spacing thickness greater than 2mm, the system does not display a pop-up message indicating the slice spacing is not optimal.

Source facts

Action
On 08/22/2024, an "URGENT: MEDICAL DEVICE CORRECTION" Letter was mailed to customers informing them of 2 software anomalies impacting spine procedures on the StealthStation S8 system (StealthStation S8 Basic, StealthStation S8 Plus and StealthStation S8 Premium) with Application software version 1.3.0 or 1.3.2. Issues are: 1.Navigation Orientation: When lateral orientation radiographic images are taken using a navigation-enabled Ziehm Vision FD Vario 3D C-arm, the orientation information appears flipped when displayed by the StealthStation S8 system. The navigation information is displayed with a 180-degree rotation around the superior/inferior(S/I) axis with the left/right (L/R) axis and anterior/posterior (A/P) axis appearing as mirror images. The exam image orientation and related indication labels on the exams remain correct. Note -this anomaly only impacts StealthStation" S8 systems where the Ziehm Vision FDVario 3D C-arm Interface Kit, 9734486, installed and calibrated on a Ziehm FD system in the lateral orientation; other orientations settings are not impacted. Medtronic Recommends Do not use the lateral orientation setting on the Ziehm Vision FD Vario 3D C-arm, if using the StealthStation" S8 with software version 1.3 or 1.3.2. 2.Pop-up Message: When using CT + Fluoroscopy images with a slice spacing thickness greater than 2 mm, the system does not display a pop-up message indicating the slice spacing is not optimal. Navigation remains accurate in this situation. Customers are instructed to: -Review Appendix A which provides detailed examples of the issues and how to identify if your system is affected. -Medtronic will work expeditiously with facilities to coordinate a software update to version 2.1 at no charge to resolve anomalies. Questions, contact Medtronic Technical Services at 1-888-826-5603 or email at rs.navtechsupport@medtronic.com.
Initiated
2024-08-22
Root cause
Software design
Distribution
Worldwide - U.S Nationwide distribution including in the states of AK, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NM, NV, NY, OH, OK, PA, Puerto Rico, SC, TN, TX, UT, VA, WA, WI and WV. The countries of Argentina, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELGIUM, Brazil, BULGARIA, CANADA, CHILE, China, Colombia, Costa Rica, CROATIA, CYPRUS, DENMARK, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, GREECE, GUADELOUPE, HUNGARY, ICELAND, India, ISRAEL, ITALY, Japan, Jordan, KAZAKHSTAN, KENYA, LEBANON, LIBYAN ARAB JAMAHIRIYA, LITHUANIA, Malaysia, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NICARAGUA, OMAN, PAKISTAN, PANAMA, Philippines, POLAND, PUERTO RICO, QATAR, Republic of Korea, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SERBIA, Singapore, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, Turkey, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM.
Recall number
Z-0050-2025
Recall status
Open, Classified
CDRH record ID
209896
FDA classified
2024-10-10
Recall event number
95283

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗