Z-0051-06: Xtra Classic Glucose Monitoring System; Recalling Firm/Manufacturer: Abbott Diabetes Care, Inc., 1360 Sou

Recall record Z-0051-06 for Abbott Laboratories Medisense Products. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0051-06
Company in FDA record
Abbott Laboratories Medisense Products
FDA classification date
Oct 21, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

User may inadvertently change the Units of Measure from mg/dL to mmol/L and the blood glucose results could be misinterpreted. This may lead to under treatment and potential for hyperglycemia.

Source facts

Action
The firm has undertaken a field correction to notify their customers of the potential for the units of measure to be changed. The firm plans to (1) include a flyer in all Precision Xtra and FreeStyle meter kits to reinforce current labeling regarding units of measure and update the Frequently Asked Questions section of the website to reinforce the units of measure change, (2) Lock out the units of measure in new production meters to eliminate potential to change the units of measure settings, update the meter kits with a flyer to override the section of the manual indicating that the units of measure are changeable, and change the product insert in glucose test strips to indicate which units of measure are appropriate for the user and (3) update the labeling to indicate that the units of measure are not user configurable to eliminate the need for the flier.
Initiated
2005-06-07
Root cause
Other
Terminated
2006-05-17
Distribution
Approximately 10,005,206 meters have been distributed worldwide since 1999, with 5,885,332 meters distributed in the US, and 4,517,211 in use.
Recall number
Z-0051-06
Recall status
Terminated
CDRH record ID
41647
FDA classified
2005-10-21
Recall event number
33301

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗