Reason stated in the source
B Braun Outlook ES Safety Infusion System "Run" Light Emitting Diodes (LEDs) pumps stop infusing and a backup alarm sounds, but the "Run" LEDs advance as if the pumps were infusing.
Source facts
- Action
- B. Braun Medical, Inc. sent an "URGENT: FIELD CORRECTION" letter dated September 7, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. A firm representative will contact all affected customers in order to schedule the field correction. The correction consists of revising the Door and Main Processor software, inspecting the main board, and replacing the board, if warranted. A Field Correction Information Form was included in the letter for customers to complete and return. Contact Technical Product Support at 1-800-627-7867 for questions regarding this notice.
- Initiated
- 2011-09-07
- Root cause
- Software design
- Terminated
- 2015-05-04
- Distribution
- Nationwide Distribution
- Recall number
- Z-0051-2012
- Recall status
- Terminated
- CDRH record ID
- 103860
- FDA classified
- 2011-10-18
- Recall event number
- 59848