Terumo Cardiovascular Systems (Terumo CVS) received one report of loss of
power to the control module and subsequently the drive motor for the Sarns
centrifugal system during cardiopulmonary bypass.
Source facts
Action
TERUMO sent an Urgent Medical Device Recall dated April 13, 2012, to all affected consignees. An Addendum was sent on September 25, 2012. All US Consignees will receive a Safety Advisory and an update to the Operator's Manual via Federal Express to inform them of the reported malfunction and potential hazards and to raise awareness of the benefits of using a retrograde valve and the importance of back-up system availability. TERUMO will include a customer response form and follow up with all affected users to obtain a response to the recall.
For questions customers should call 1-800-521-2818.
For questions regarding this recall call 734-741-6173.
Initiated
2012-09-24
Root cause
Component design/selection
Terminated
2013-11-13
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Bangladesh, Seoul Korea, Chile, Brazil, Venesuela, Thailand, Mexico, India, Taiwan, Canada, Dominican Republic, Guatemala, Ecuador, Germany, Republic of Georgia, Hong Kong, Japan, United Arab Emirates (UAE), Japan, Belgium, Phillipines, Singapore, and South Africa.
Recall number
Z-0051-2013
Recall status
Terminated
CDRH record ID
109800
FDA classified
2012-10-11
Recall event number
62059