Open original FDA record ↗Recall record
- FDA identifier
- Z-0052-2014
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- Oct 23, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Software issue results in inaccurate 3D image may result in ablation in wrong location.
Source facts
- Action
- Siemens sent an Customer Safety Advisory Notification letter to all affected customers on July 10, 2013.The letter identified the product the problem and the action needed to be taken by the customer Letters provide steps to mitigate the risk and inform users that a software fix will be installed as soon as it is available. Siemens Service Engineers will contact all locations to arrange installation once fix is available. All personnel who use the ACUSON should be notified of the issue. Update September 17, 2013 : After discussion with CDRH, Siemens revised their letter to state that procedures using the ACUSON SC2000 system with BioSense Webster CARTO 3 equipped with CARTOSOUND Module and SoundSTAR Catheter should not be performed until the users system has been updated with the software upgrade that resolves the issue. For further questions please call (650) 694-5398.
- Initiated
- 2013-07-10
- Root cause
- Device Design
- Terminated
- 2013-11-25
- Distribution
- Worldwide Distribution - US (nationwide) including the states of IL, PA, AZ, OR, OH, NY, CO, TX and CA., and the countries of France, Germany, Japan, Portugal, Saudi Arabia, Taiwan and Turkey.
- Recall number
- Z-0052-2014
- Recall status
- Terminated
- CDRH record ID
- 120286
- FDA classified
- 2013-10-23
- Recall event number
- 65812