In some systems, the unit continues to emit RF energy after the PAUSE button is pressed.
Source facts
Action
Consignees were sent a recall notification by mail dated 07/13/2017. Customers were notified of the reason for recall and advised that the PAUSE button not be used and the STOP button be used instead to stop the RF emission. Additionally, units are to be connected to the internet so that the BIOS service team can access the units to provide a software update to correct the issue. If a unit cannot be connected to the internet, customers are to contact Thermi to schedule an update.
Initiated
2017-07-13
Root cause
Software design
Terminated
2020-09-15
Distribution
AZ, AR, CA, FL, GA, ID, IL, KS, LA, MA, MI, MS, MO, NV, NJ, NM, NY, OH, OR, PA, SD, TX, WA, WI, and WV
Brazil, Canada, Israel, Japan, & Thailand.
Recall number
Z-0052-2018
Recall status
Terminated
CDRH record ID
157831
FDA record created
2017-10-27
Recall event number
77754