Skip to content

Public FDA record

Z-0052-2018: THERMI250 Radiofrequency System. Intended to provide heating for the purpose of elevating tissue tempera

Recall record Z-0052-2018 for Bios S.r.l.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0052-2018
Company in FDA record
Bios S.r.l.
FDA record creation date
Oct 27, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In some systems, the unit continues to emit RF energy after the PAUSE button is pressed.

Source facts

Action
Consignees were sent a recall notification by mail dated 07/13/2017. Customers were notified of the reason for recall and advised that the PAUSE button not be used and the STOP button be used instead to stop the RF emission. Additionally, units are to be connected to the internet so that the BIOS service team can access the units to provide a software update to correct the issue. If a unit cannot be connected to the internet, customers are to contact Thermi to schedule an update.
Initiated
2017-07-13
Root cause
Software design
Terminated
2020-09-15
Distribution
AZ, AR, CA, FL, GA, ID, IL, KS, LA, MA, MI, MS, MO, NV, NJ, NM, NY, OH, OR, PA, SD, TX, WA, WI, and WV Brazil, Canada, Israel, Japan, & Thailand.
Recall number
Z-0052-2018
Recall status
Terminated
CDRH record ID
157831
FDA record created
2017-10-27
Recall event number
77754

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗