Open original FDA record ↗Recall record
- FDA identifier
- Z-0053-04
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- Oct 24, 2003
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The detector's radial brake may not be applied properly, allowing unintended motion which could result in injury to the patient or operator.
Source facts
- Action
- Siemens sent Important Safety Customer Advisory letters dated 8/27/03 to all e.cam customers on 9/5-12/03. The accounts were informed of the problem with the detector''s radial brake not being applied properly, allowing slow, unintended downward motion of the detector that could result in injury to the patient or operator, and were advised not to leave the patient unattended while on the pallet and not to leave the patient under the detector upon completion of the study. The customers were advised that they would be contacted by their Siemens service provider to schedule a mandatory update to the e.cam to correct the problem. The upgrade will be done free of charge.
- Initiated
- 2003-09-05
- Root cause
- Other
- Terminated
- 2005-02-24
- Distribution
- Nationwide, United Arab Emirates, Albania, Argentina, Austria, Australia, Bosnia, Barbados, Bangladesh, Belgium, Bolivia, Brazil, Cameroon, Canada, Chile, China, Colombia, Republic of Congo, Costa Rica, Czech Republic, Denmark, Dominican Republic, El Salvador, Ecuador, Egypt, France, Finland, Germany, Great Britain, Georgia, Greece, Hong Kong, Honduras, Ireland, India, Italy, Iran, Iraq, Jordan, Japan, Korea, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Mexico, Morocco, Macedonia, Madagascar, Myanmar, Mauritius, Malaysia, Nepal, Norway, New Zealand, Oman, Peru, Philippines, Pakistan, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Turkey, Ukraine, Uzbekistan, Zimbabwe, South Africa and the Netherlands.
- Recall number
- Z-0053-04
- Recall status
- Terminated
- CDRH record ID
- 29204
- FDA classified
- 2003-10-24
- Recall event number
- 27105