Reason stated in the source
System Warning Alarm failure: When a patient being monitored is in a pre-existing condition of continuous MESSAGE or ADVISORY level alarm preceding a SYSTEM WARNING level alarm, the SYSTEM WARNING audible alarm and flashing yellow border around the patient panel at the CIC does not occur.
Source facts
- Action
- An Urgent Medical Device Correction letter, dated June 23, 2006, was sent to customers notifying them of the issue, provided short term recommendations and provided the long term solution as a software upgrade that will correct this situation. A reply card is asked to be returned to initiate the upgrade process.
- Initiated
- 2006-06-23
- Root cause
- Other
- Terminated
- 2007-05-25
- Distribution
- Nationwide distribution --- including states of AL, AR, AZ, CA, CO, Washington DC, FL, IL, IN, ME, MI, MN, MO, NC, NE, NJ, NY, OK, PA, SC, TX, VA, WA, WI, WV.