No results/incorrect results due to failure of the hybridization heater.
Source facts
Action
On or about 11/19/2018 there was one complaint from a customer regarding a thermal control failure. In response to the complaint the Field Service Engineer replaced hybridization heater and ran all tests as passed. Subsequently, the customer reported the initial service was ineffective. The VERIGENE SP Processor was then removed from the customer site and returned for analysis.
No Customer Notification was communicated for this removal.
Initiated
2018-12-10
Root cause
Component design/selection
Distribution
US distribution in MO.
No OUS
Recall number
Z-0054-2021
Recall status
Open, Classified
CDRH record ID
183478
FDA record created
2020-10-05
Recall event number
86358