Open original FDA record ↗Recall record
- FDA identifier
- Z-0056-2021
- Company in FDA record
- Brainlab AG
- FDA record creation date
- Oct 6, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Brainlab Ultrasound Navigation Software does not support the modification of the probe's image width.
Source facts
- Action
- The intraoperative Brainlab Ultrasound Navigation Software does not support the modification of the probes image width, an optional feature of the BK Medical Ultrasound System, resulting in an incorrect display of the depth of the ultrasound overlay by the navigation for non-linear ultrasound probes, with the deviation depending on the probe and the difference between the calibrated and used image width.
- Initiated
- 2020-08-31
- Root cause
- Software design
- Terminated
- 2021-02-17
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of Illinois, Massachusetts, Texas, California, Idaho, Missouri, Oregon, Connecticut and the countries of Australia, Austria, Belgium, Brazil, Canada, China, Curacao, Czech Republic, Egypt, Finland, France, Germany, Switzerland, Netherlands, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kuwait, Malaysia, Myanmar, Norway, Pakistan, Romania, Slovakia, South Africa, Spain, Sweden, Thailand, Turkey and United Kingdom.
- Recall number
- Z-0056-2021
- Recall status
- Terminated
- CDRH record ID
- 183720
- FDA record created
- 2020-10-06
- Recall event number
- 86438