Open original FDA record ↗Recall record
- FDA identifier
- Z-0056-2025
- Company in FDA record
- Olympus Corporation of the Americas
- FDA classification date
- Oct 11, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Certain serial numbers of the CV-190 do not start up properly because parts that deviated from the specification were assembled into the power supply unit.
Source facts
- Action
- An URGENT: MEDICAL DEVICE FIELD CORRECTIVE ACTION notification letter dated 9/14/23 was sent to customers. Action steps to be taken by the end user: Olympus has determined based upon our distribution records that your facility is in possession of one or more affected CV-190 with a serial number shown in Attachment 1. Olympus requires that you take the following actions: 1. Inspect your inventory for the referenced products and identify any devices with the model number and serial number specified in Attachment 1. Cease use of products and quarantine any affected products. Please check all areas of the hospital to determine if any of these products remain in inventory. 2. Please contact our Olympus Repair Center at 1-800-848-9024, option 3, to arrange for a service to exchange the power supply unit in affected devices. Olympus regrets any inconvenience caused and fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at 647-999-3203 or Cynthia.Ow@Olympus.com.
- Initiated
- 2023-09-14
- Root cause
- Nonconforming Material/Component
- Distribution
- US Nationwide distribution.
- Recall number
- Z-0056-2025
- Recall status
- Open, Classified
- CDRH record ID
- 203683
- FDA classified
- 2024-10-11
- Recall event number
- 93217