Z-0056-2025: EVIS EXERA III Video System Center Model Number CV-190. Intended to be used with OLYMPUS camera heads, en

Recall record Z-0056-2025 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0056-2025
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Oct 11, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Certain serial numbers of the CV-190 do not start up properly because parts that deviated from the specification were assembled into the power supply unit.

Source facts

Action
An URGENT: MEDICAL DEVICE FIELD CORRECTIVE ACTION notification letter dated 9/14/23 was sent to customers. Action steps to be taken by the end user: Olympus has determined based upon our distribution records that your facility is in possession of one or more affected CV-190 with a serial number shown in Attachment 1. Olympus requires that you take the following actions: 1. Inspect your inventory for the referenced products and identify any devices with the model number and serial number specified in Attachment 1. Cease use of products and quarantine any affected products. Please check all areas of the hospital to determine if any of these products remain in inventory. 2. Please contact our Olympus Repair Center at 1-800-848-9024, option 3, to arrange for a service to exchange the power supply unit in affected devices. Olympus regrets any inconvenience caused and fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at 647-999-3203 or Cynthia.Ow@Olympus.com.
Initiated
2023-09-14
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution.
Recall number
Z-0056-2025
Recall status
Open, Classified
CDRH record ID
203683
FDA classified
2024-10-11
Recall event number
93217

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗