Open original FDA record ↗Recall record
- FDA identifier
- Z-0057-04
- Company in FDA record
- Hill-Rom Manufacturing, Inc.
- FDA classification date
- Oct 24, 2003
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Power cord overheating
Source facts
- Action
- Consignees were notified by letter sent January 14, 2002 warning them of the hazard involved with the defective part. They were informed that Hill Rom needed to replace the power cord on each unit to correct the possibility of the power cord from overheating, melting or smoking. The maintenance would be performed at the location of each unit (most of which are at their rental facilities located nationwide) by the firm''s servicing representatives. Meanwhile, customers were urged to check the power cord to ensure that it is fully inserted into the wall outlet and the connector on the Clinitron At Home Air Fluidized Therapy Unit. The modification was to be completed by August 31, 2002.
- Initiated
- 2002-01-14
- Root cause
- Other
- Terminated
- 2003-10-27
- Distribution
- Hill Rom primarily leases these units to home care patients and some are sold nationwide.
- Recall number
- Z-0057-04
- Recall status
- Terminated
- CDRH record ID
- 29401
- FDA classified
- 2003-10-24
- Recall event number
- 27190