Skip to content

Public FDA record

Z-0057-04: Clinitron At Home Air Fluidized Therapy Unit

Recall record Z-0057-04 for Hill-Rom Manufacturing, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0057-04
Company in FDA record
Hill-Rom Manufacturing, Inc.
FDA classification date
Oct 24, 2003
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Power cord overheating

Source facts

Action
Consignees were notified by letter sent January 14, 2002 warning them of the hazard involved with the defective part. They were informed that Hill Rom needed to replace the power cord on each unit to correct the possibility of the power cord from overheating, melting or smoking. The maintenance would be performed at the location of each unit (most of which are at their rental facilities located nationwide) by the firm''s servicing representatives. Meanwhile, customers were urged to check the power cord to ensure that it is fully inserted into the wall outlet and the connector on the Clinitron At Home Air Fluidized Therapy Unit. The modification was to be completed by August 31, 2002.
Initiated
2002-01-14
Root cause
Other
Terminated
2003-10-27
Distribution
Hill Rom primarily leases these units to home care patients and some are sold nationwide.
Recall number
Z-0057-04
Recall status
Terminated
CDRH record ID
29401
FDA classified
2003-10-24
Recall event number
27190

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗