Reason stated in the source
Misassembly: The ON-Q PainBuster PMB01 package may contain a Soaker catheter instead of a Non-Soaker catheter. As a result, the bolus button may remain latched in the down position which may cause the the device to deliver continuously at 7 ml/hr instead of 2 ml/hr.
Source facts
- Action
- An Important Recall Notification Letter was faxed to affected customers on March 27, 2007. Affected customers were notified of the recall, the reason for recall, and instructed to immediately quarantine any remaining inventory. A Customer Response (and Fax back) form was included with the Recall Notification. Affected customers were requested to return the FAX back form to I-Flow within 48 hours of receipt. Replacement product will be sent out as applicable.
- Initiated
- 2007-03-27
- Root cause
- Process control
- Terminated
- 2009-10-07
- Distribution
- Worldwide, USA and Australia
- Recall number
- Z-0057-2008
- Recall status
- Terminated
- CDRH record ID
- 64546
- FDA classified
- 2007-10-06
- Recall event number
- 44933