Open original FDA record ↗Recall record
- FDA identifier
- Z-0058-2013
- Company in FDA record
- Belmont Instrument Corporation
- FDA classification date
- Oct 12, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Leak from a saline bag caused saline to drip along the back of the Rapid infuser close to the Power Entry Module causing an electrical flash at the power entry area.
Source facts
- Action
- Belmont Instrument initiated recall on 9/25/12. A "Field Correction" notice was provided that described the product issue and recommended actions. US-based sales representatives visited all facilities to provide the modified power cord/shield. For units sold outside the United States, the modified power cords were sent to distributors who would ensure the new power cords were provided to all facilities within their accounts.
- Initiated
- 2012-09-25
- Root cause
- Device Design
- Terminated
- 2014-02-28
- Distribution
- Worldwide Distribution including Nationwide (USA) and the countries of Australia, Austria, Belgium, Brazil, Canada, China, Denmark, England, Estonia, France, Germany, Hong Kong, India, Israel, Italy, Jordan, Mongolia, Netherlands, New Zealand, Norway, Pakistan, Peru, Puerto Rico, S. Korea, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Turkey, UK, United Arab Emirates, and Vietnam.
- Recall number
- Z-0058-2013
- Recall status
- Terminated
- CDRH record ID
- 111447
- FDA classified
- 2012-10-12
- Recall event number
- 62703