Skip to content

Public FDA record

Z-0059-2008: Philips Medical System - BV Pulsera Mobile C-arms with Remote Control Option, (Release 2)

Recall record Z-0059-2008 for Philips Medical Systems North America Co. Phillips. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0059-2008
Company in FDA record
Philips Medical Systems North America Co. Phillips
FDA classification date
Nov 10, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

System Shut Down -- Unexpected Shut Down during use of the Optional Remote Control.

Source facts

Action
On 8/23/07, the firm sent out a Safety Notification letter, via certified mail. On 8/29/07, the firm sent out an URGENT - DEVICE CORRECTION letter via regular mail, dated 8/20/07. Both letters advised customers that if the device shuts down during use, it can be restarted within about 1 minute by switching the system off and then on again, using the system on/off buttons. The letter(s) states that software is being developed to correct the problem.
Initiated
2007-08-23
Root cause
Software design
Terminated
2010-09-27
Distribution
Nationwide
Recall number
Z-0059-2008
Recall status
Terminated
CDRH record ID
55372
FDA classified
2007-11-10
Recall event number
39392

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗