Z-0059-2021: 1500 Flex Heat and Moisture Exchanger, Catalog Reference 150 - Product Usage: Heat Moisture Exchangers (

Recall record Z-0059-2021 for AB ULAX. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0059-2021
Company in FDA record
AB ULAX
FDA record creation date
Oct 6, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Firm has received 6 complaints that the Heat Moisture Exchanger (HME) disconnected from the closed respiratory circuit during use. If the HME separates, the ventilator circuit may not provide adequate ventilation to the patient, potentially resulting in lack of oxygen leading to hypoxia or death.

Source facts

Action
The firm distributed customer letters dated September 4, 2020, which informed customers of the product issue. Distributors were asked to inform customers of the recall. End users were asked to do the following: Inventory that is Currently In-Use: 1. Please evaluate any currently ventilated patient who may have such a device (any BALLARD Heat Moisture Exchanger (HME) 150) in their airway circuit, to determine if they belong to any of the impacted product lots. 2. If any of your patient(s) do have a potentially impacted device in their airway circuit, you must replace such device with a commercially available alternative as soon as possible. Due to the potential for brain damage or death if a circuit interruption were not noticed for about 4 minutes, this is a critical instruction. If, despite this instruction and at your own discretion , you elect still not to do so, then you must check carefully to ensure that both ends of the device are seated very firmly within the circuit; consider taping the device in place at both ends to avoid any potential for disconnections; and ensure that all alarms are working, monitored and addressed immediately if indicated. 3. Please complete the attached End-User Product Recall Response Form and return a copy of the form by email to: Avanos3250@stericycle.com or fax: (888) 229-0320. 4. Once the form is received, a representative will contact you to facilitate return of impacted BALLARD Heat Moisture Exchanger (HME) 150 . 5. Avanos will manage the return of the impacted BALLARD Heat Moisture Exchanger (HME) 150. Please respond within five (5) business days of receipt of this letter. Inventory Already Consumed 6. Please complete the attached End-User Product Recall Response Form (Attachment 1) and select the appropriate box to indicate that you have no inventory of impacted products. 7. Please return this form to Avanos by either email: Avanos3250@stericycle.com or fax: (888)
Initiated
2020-09-04
Root cause
Process control
Terminated
2023-02-10
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Sweden, Taiwan, Hong Kong, and Spain.
Recall number
Z-0059-2021
Recall status
Terminated
CDRH record ID
183668
FDA record created
2020-10-06
Recall event number
86417

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗