Philips Medical Systems North America Co. Phillips
FDA classification date
Nov 10, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
System Shut Down -- Unexpected Shut Down during use of the Optional Remote Control.
Source facts
Action
On 8/23/07, the firm sent out a Safety Notification letter, via certified mail. On 8/29/07, the firm sent out an URGENT - DEVICE CORRECTION letter via regular mail, dated 8/20/07. Both letters advised customers that if the device shuts down during use, it can be restarted within about 1 minute by switching the system off and then on again, using the system on/off buttons. The letter(s) states that software is being developed to correct the problem.
Initiated
2007-08-23
Root cause
Software design
Terminated
2010-09-27
Distribution
Nationwide
Recall number
Z-0060-2008
Recall status
Terminated
CDRH record ID
55373
FDA classified
2007-11-10
Recall event number
39392