Skip to content

Public FDA record

Z-0060-2021: DJO EMPOWR Tibial Knee Punch Handle Part No. 801-05-040 and 801-05-018 contained in Instrument Tray for E

Recall record Z-0060-2021 for Encore Medical, LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0060-2021
Company in FDA record
Encore Medical, LP
FDA classification date
Oct 7, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for the retaining pin of the tibial punch handle to become loose and fall out while impacting. If undetected, there is a potential of leaving debris in the joint space.

Source facts

Action
On August 3, 2020, Field Safety Notices were sent to the firm's distributors by email advising them of this field action, informing them that the firm has received two product complaints where an issue has been experienced with the retaining pin becoming loose and falling out while impacting. Distributors were informed that no adverse events have been reported but that there is a potential of leaving debris in the joint space if this issue were to occur and remain undetected. The letter requests distributors to communicate the following information to implanting surgeons to avoid the potential of leaving debris in the joint space: 1. Sales representative is to inspect instrumentation before every surgery per the Instructions For Use. A small object should be pushed against the end of the pin to confirm it is securely fixed in place. 2. The sales rep should warn of the possibility of the loose pin and steps to take to confirm containment of the loose pin in order to mitigate patient risk. 3. After use of the instrument, it should be verified that the pin remains fixed in place. Distributors were informed that when devices that meet specification become available for the 5 impacted lots, a field exchange will occur as part of a reportable recall. Distributors are requested to complete the acknowledgment form and return it by 8-10-2020 to confirm that they have read and understood best practices for use of the knee punch handle, and have disseminated the information to all employees and contracted Sales Representatives. Customers with questions should contact Teffany Hutto by phone at (512) 834-6255 or by email at teffany.hutto@djoglobal.com. On 03/12/2021, recall notices were emailed to additional customers informing them of additional affected models/lots.
Initiated
2020-08-03
Root cause
Process change control
Distribution
U.S. Nationwide distribution including in the states of VA, TN, MN, TX, IL, IN, KY, OH, SC, RI, MD, PA, NY, CA, KS, OK, AR, LA, FL, AL, GA, AZ, NV, WA, MS, ME, NV and Puerto Rico.
Recall number
Z-0060-2021
Recall status
Open, Classified
CDRH record ID
183234
FDA classified
2020-10-07
Recall event number
86257

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗