Open original FDA record ↗Recall record
- FDA identifier
- Z-0061-06
- Company in FDA record
- Baxter Healthcare Corp.
- FDA classification date
- Oct 15, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
If an administrative user modifies any of the seven pre-defined Monitoring Parameters when they are attached to a medication in the drug file, a Point of Care end user will be unable to enter monitoring parameter results on the handheld.
Source facts
- Action
- Urgent Device Correction recall letters dated 8/29/05 were sent to the direct accounts on the same date via first class mail. The letter describes the issue with the modification or creation of a Monitoring Parameter, and informed the accounts that modifying the Monitoring Parameters could result in a potential for a patient receiving an incorrect medication or an incorrect medication dosage. The accounts were advised that administrative end users should never modify the seven pre-defined Monitoring Parameters or create new Monitoring Parameters.
- Initiated
- 2005-08-29
- Root cause
- Other
- Terminated
- 2008-02-06
- Distribution
- Missouri, Massachusetts, Vermont, California and Canada.
- Recall number
- Z-0061-06
- Recall status
- Terminated
- CDRH record ID
- 41544
- FDA classified
- 2005-10-15
- Recall event number
- 33250