Open original FDA record ↗Recall record
- FDA identifier
- Z-0061-2008
- Company in FDA record
- Zoll Lifecor Corporation
- FDA classification date
- Oct 18, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Possible malfunction of the battery pack. A significant electrostatic discharge (ESD) can cause the battery pack to malfunction. The malfunction causes the battery pack to discharge more rapidly and prevents the pack from recharging.
Source facts
- Action
- Patients currently using the LifeVest should continue to do so while the product removal is underway. Advisory notices will be sent via first class mail 9/14/2007 to all active patients using the LifeVest device. Product replacements began on 8/31/07 by shipping out revised battery packs for new patient use. Shipments of new battery packs will continue based on production capacity and material availability. Any patients expected to be using the device longer than 5 months will be provided with 2 replacement Battery Packs. They will be instructed to return their existing packs in the same shipping box.
- Initiated
- 2007-09-14
- Root cause
- Device Design
- Terminated
- 2008-06-13
- Distribution
- The product was shipped nationwide to patients. The product was also shipped to distributors in Germany, United Kingdom, Italy, Spain, France, and Denmark.
- Recall number
- Z-0061-2008
- Recall status
- Terminated
- CDRH record ID
- 64559
- FDA classified
- 2007-10-18
- Recall event number
- 44942