Open original FDA record ↗Recall record
- FDA identifier
- Z-0061-2010
- Company in FDA record
- Western - a Scott Fetzer Company
- FDA classification date
- Oct 15, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The yoke pins of the attached medical gas yoke assemblies could become displaced to such a degree that the yoke pin index safety system could become compromised and allow a post valve of one gas type to become interchangeable with a yoke connection of another gas type.
Source facts
- Action
- Western Enterprises issued an "Urgent-Device Recall" notification dated August 20, 2009 informing Consignees of the affected product. Consignees are to contact the firm to request backup plates and installation instructions or request an RGA and shipping instructions for return of the manifold block assembly for modification. For further information, contact Western at 1-800-783-7890.
- Initiated
- 2009-08-10
- Root cause
- Component design/selection
- Terminated
- 2012-04-25
- Distribution
- Nationwide Distribution -- AL, AR, CA, CO, FL, GA, HI, IL, IN, MA, OH, OR, PA, TX, WA and WI.
- Recall number
- Z-0061-2010
- Recall status
- Terminated
- CDRH record ID
- 84136
- FDA classified
- 2009-10-15
- Recall event number
- 52847