Z-0062-05: Infusion Dynamics Power Infuser pump. Model number M100B-3A
Recall record Z-0062-05 for Infusion Dynamics, A Div. of Zoll Med Corpinfusion Dynamics. View source facts, labeled dates, and the original FDA record.
Infusion Dynamics, A Div. of Zoll Med Corpinfusion Dynamics
FDA classification date
Nov 2, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
Front Panel may be cracked on some units. If cracked fluid can enter the infuser and short the start/stop button. Shorting would cause the infuser to stay in the open position.
Source facts
Action
The recalling firm issued a recall letter dated 9/20/04 to their customers informing them of the problem and the need to return the units.
Initiated
2004-09-21
Root cause
Other
Terminated
2005-02-07
Distribution
The products were shipped to dealers in TX, MO, and VT; medical facilities in CA, NV, and NY. The product was also shipped to government accounts in PA and Germany.
Recall number
Z-0062-05
Recall status
Terminated
CDRH record ID
35298
FDA classified
2004-11-02
Recall event number
30168