Open original FDA record ↗Recall record
- FDA identifier
- Z-0062-06
- Company in FDA record
- Event Medical
- FDA classification date
- Oct 15, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Firm received reports of combustion within the blender system of the Inspiration Ventilator.
Source facts
- Action
- Customer shall be notified by an Urgent Product Safety Notfication letter dated August 12, 2005. The letter shall advise the customer with the specifics of this problem, summarise the fact that the hazard potential in negligible and will advise the customer regarding options for replacement of the Blender Module assembly (part number F710400) and the Air/O2 Filters (part number F910300). The letter will also provide the customers with a set of actions which can be completed immediately in order to ensure the integrity of their gas supplies until their unit has been upgraded.
- Initiated
- 2005-08-12
- Root cause
- Other
- Terminated
- 2012-02-29
- Distribution
- Nationwide and Canada, Europe, Japan, Latin America, Asia/Pacific, Middle East
- Recall number
- Z-0062-06
- Recall status
- Terminated
- CDRH record ID
- 41591
- FDA classified
- 2005-10-15
- Recall event number
- 33276