Open original FDA record ↗Recall record
- FDA identifier
- Z-0062-2017
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA record creation date
- Oct 8, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The detectors on the Symbia S and Symbia T camera systems are supported and moved horizontally and vertically by four linear bearing cars. There is a potential for the linear bearing car to fail allowing the ball bearings to fall out.
Source facts
- Action
- Customers were contacted by a service organization that performed product inspections at their sites.
- Initiated
- 2013-08-16
- Root cause
- Component design/selection
- Terminated
- 2016-12-23
- Distribution
- Worldwide Distribution - US: Nationwide, including Washington D.C., and Puerto Rico. and the countries of: Poland, France, Austria, Czech Republic, Ireland, Germany, Italy, Japan, Canada, United Arab Emirates, Turkey, Netherlands, United Kingdom, Brazil, South Africa, Kuwait, Belgium, Russia, Taiwan, Australia, Bangladesh, Bulgaria, Brazil, Switzerland, Chile, China, Colombia, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Finland, Hong Kong, Croatia, Hungary, Indonesia, Iran, Israel, India, South Korea, Kuwait, Kazakhstan, Lithuania, Latvia, Morocco, Mexico, Norway, New Zealand, Peru, Phillipines, Pakistan, Portugal, Qatar, Romania, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Thailand, Tunisia, Taiwan, Venezuela, and Vietnam.
- Recall number
- Z-0062-2017
- Recall status
- Terminated
- CDRH record ID
- 122574
- FDA record created
- 2016-10-08
- Recall event number
- 66544