Z-0062-2019: Laparoscopy Instruments Instructions For Use (IFU) for the following: 1)Monopolar electrode, 2) Insert, h

Recall record Z-0062-2019 for Dannoritzer Medizintechnik GmbH & Co. KG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0062-2019
Company in FDA record
Dannoritzer Medizintechnik GmbH & Co. KG
FDA record creation date
Oct 3, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

The Manual cleaning process and the usability of the reprocessing instruction by the user had not been validated. The IFU was modified to remove the manual cleaning method and revised the instructions for use for clarity and readability.

Source facts

Action
In July 2017, Dannoritzer Medizintechnik GmbH & Co. KG issued emails to customers notifying them that an updated Instructions For Use (IFU) manual would be distributed by July 27, 2018. On August 8, 2018, Dannoritzer Medizintechnik GmbH & Co. KG issued Urgent Medical Device Field Correction notices dated August 6, 2018 to customers along with an updated IFU. Customers were encouraged to immediately remove from their place of use and destroy the existing original DANmed Laparoscopy Instruments IFU as well as any copies and replace them with the revised version (V.2018-08-03) included with the notice. Distributors or those who have further distributed the product, please provide those customers with a copy of the notice and revised IFU. Customers are also encouraged to complete and return the Confirmation Card and email to notification@danrepair.de Customers with questions may contact Stefan Rauch 011-49-7461-9623915 8 am - 12 PM, EST or email notification@danrepair.de
Initiated
2018-08-06
Root cause
Labeling False and Misleading
Terminated
2019-07-22
Distribution
US Nationwide in the States of OH, MI, MA, CA, OK, TN, GA, FL, IL, NC, TX, MO.
Recall number
Z-0062-2019
Recall status
Terminated
CDRH record ID
167152
FDA record created
2018-10-03
Recall event number
80970

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗