The Manual cleaning process and the usability of the reprocessing instruction by the user had not been validated. The IFU was modified to remove the manual cleaning method and revised the instructions for use for clarity and readability.
Source facts
Action
In July 2017, Dannoritzer Medizintechnik GmbH & Co. KG issued emails to customers notifying them that an updated Instructions For Use (IFU) manual would be distributed by July 27, 2018. On August 8, 2018, Dannoritzer Medizintechnik GmbH & Co. KG issued Urgent Medical Device Field Correction notices dated August 6, 2018 to customers along with an updated IFU.
Customers were encouraged to immediately remove from their place of use and destroy the existing original DANmed Laparoscopy Instruments IFU as well as any copies and replace them with the revised version (V.2018-08-03) included with the notice.
Distributors or those who have further distributed the product, please provide those customers with a copy of the notice and revised IFU.
Customers are also encouraged to complete and return the Confirmation Card and email to notification@danrepair.de
Customers with questions may contact Stefan Rauch 011-49-7461-9623915 8 am - 12 PM, EST or email notification@danrepair.de
Initiated
2018-08-06
Root cause
Labeling False and Misleading
Terminated
2019-07-22
Distribution
US Nationwide in the States of OH, MI, MA, CA, OK, TN, GA, FL, IL, NC, TX, MO.
Recall number
Z-0062-2019
Recall status
Terminated
CDRH record ID
167152
FDA record created
2018-10-03
Recall event number
80970