Z-0063-2008: Medtronic Export XT Aspiration Catheter, 6F (US) Medtronic Export XT Aspiration Catheter, CE, 6FR (Outsid

Recall record Z-0063-2008 for Medtronic, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0063-2008
Company in FDA record
Medtronic, Inc.
FDA classification date
Oct 31, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Additional Warning in Instructions for Use for guide wire management. Complaints of wire lumen tearing during use or band marker becoming dislodged.

Source facts

Action
Medtronic has notifed sales representatives to visit Cath lab site with the letter dated August 29, 2007 and educate the user, provide the Safety Alert Letter and document delivery of the letter with the confirmation form.
Initiated
2007-08-29
Root cause
Labeling design
Terminated
2008-09-08
Distribution
Worldwide, including USA, Canada, India, Austria, Belgium, France, Germany, Italy, Netherlands, Spain, Sweden, Switzerland, and UK.
Recall number
Z-0063-2008
Recall status
Terminated
CDRH record ID
64325
FDA classified
2007-10-31
Recall event number
44724

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗