Open original FDA record ↗Recall record
- FDA identifier
- Z-0063-2018
- Company in FDA record
- North Star Imaging Inc
- FDA record creation date
- Nov 6, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
It was discovered a low-level radiation leak that was found on the NSI X-7000 system.
Source facts
- Action
- North Star Imaging will bring defect into compliance by the following : The Center for Devices and Radiological Health (CDRH) understands that the CAP involving 3 units will consist of the following: 1. All three customers have been notified that the affected products will be repaired free of charge. 2 .The area where the emission rate could be exceeded will have shielding added until emissions are adequately attenuated under the conditions described above. 3. The repaired systems will be completely surveyed to verify compliance with the cabinet x- ray system performance standards emission limit. 4.All repairs will be completed before February 1, 2018. For further questions, please call (800) 635-8392.
- Initiated
- 2017-08-23
- Root cause
- Radiation Control for Health and Safety Act
- Distribution
- US Distribution to the states of : PA, OH and KS.
- Recall number
- Z-0063-2018
- Recall status
- Open, Classified
- CDRH record ID
- 159021
- FDA record created
- 2017-11-06
- Recall event number
- 78254