Skip to content

Public FDA record

Z-0063-2025: 1.2mm Cross Cut Fissure - Intended for use in the cutting, drilling, reaming, decorticating, shaping and

Recall record Z-0063-2025 for Stryker Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0063-2025
Company in FDA record
Stryker Corporation
FDA classification date
Oct 11, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

May exhibit endotoxin levels greater than 2.15 EU/device, fever and inflammation may occur from endotoxin exposure and meningitis may only occur if there is contact with cerebrospinal fluid and thus, neurosurgical/spinal procedures may be considered a population at greatest risk.

Source facts

Action
Stryker issued Urgent Medical Device Recall Letter on 9/5/24 via FedEx 2-day to Recall Department, Materials Manager, OR Director, Risk Manager. Letter states reason for recall, health risk and action to take: Actions needed 1. Immediately review your inventory to locate and quarantine any affected products at your facility. 2. Return the enclosed Business Reply Form (BRF), even if the affected product is no longer in inventory, to the email Instruments.Recalls@stryker.com. 3. Upon receipt of the completed BRF, Stryker will provide a shipping label to return recalled product on-hand. 4. A replacement will be provided upon receipt of the recalled product. 5. Maintain awareness of this communication internally and inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients accordingly. If you have any questions about this matter, contact Stryker Instruments recall team directly by calling (269) 568-1481 or email at Instruments.Recalls@stryker.com.
Initiated
2024-09-05
Root cause
Process control
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Columbia.
Recall number
Z-0063-2025
Recall status
Open, Classified
CDRH record ID
210037
FDA classified
2024-10-11
Recall event number
95342

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗