Open original FDA record ↗Recall record
- FDA identifier
- Z-0064-2015
- Company in FDA record
- Roche Diagnostics Operations, Inc.
- FDA classification date
- Oct 10, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
The default references for normal values are inconsistent between cobas b 221 and other blood gas instruments. In addition, the reference ranges are different when comparing with corresponding ranges listed in the Instructions for Use for the analyzer and the reference literature source.
Source facts
- Action
- The Urgent Medical Device Correction will be sent via UPS Ground (receipt required) to the consignee list. The Urgent Medical Device Correction will be sent via UPS Ground (receipt required) to the consignee list. Consignees are instructed as follows: " Check the current set of reference values on your cobas b 221 blood gas system and, if appropriate, change the reference ranges accordingly. " If you have not established your own reference ranges, please follow the instructions in the Workaround section of the UMDC or contact the Roche Support Network Customer Support Center.
- Initiated
- 2014-09-12
- Root cause
- Labeling Change Control
- Terminated
- 2018-01-10
- Distribution
- Nationwide Distribution including Puerto Rico and Guam, and the states of AK, AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, ME, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA and WI.
- Recall number
- Z-0064-2015
- Recall status
- Terminated
- CDRH record ID
- 129802
- FDA classified
- 2014-10-10
- Recall event number
- 69173