Reason stated in the source
These devices have an incorrect specification against safety standard ISO61010-1. Should a single fault error occur, a mains wire can contact a metal plate which is accessible to the user. This could lead to mains power on the metal plate resulting in hazardous voltage to the user.
Source facts
- Action
- On approximately 06/13/2018 sent a Field Action Notices to consignees. Instructions included to make arrangements with their Leica Service Engineer to correct affected devices, ensure all appropriate personnel are made aware of the recall, and complete and return the Field Action Notice Return Response Form to confirm receipt and understanding of the recall notice.
- Initiated
- 2018-06-13
- Root cause
- Device Design
- Terminated
- 2019-03-18
- Distribution
- Distributed nationwide to AZ, CA, CO, KS, MA, MH, MO, NJ, OH, WI. Distributed worldwide to CANADA, CHINA, GERMANY, GREAT BRITAIN, JAPAN, SWITZERLAND
- Recall number
- Z-0064-2019
- Recall status
- Terminated
- CDRH record ID
- 166939
- FDA record created
- 2018-10-03
- Recall event number
- 80916