Open original FDA record ↗Recall record
- FDA identifier
- Z-0065-2017
- Company in FDA record
- Winco Mfg., LLC
- FDA record creation date
- Oct 10, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Non-conforming component caused arms of TMM4 Stretcher chair to no longer meet design specifications.
Source facts
- Action
- Customers were sent an Urgent: Medical Device Recall notification on 7/12/2016 for the specific models of the TransMotion (TMM) Stretcher chairs. The letter explained the reason for the recall and asked customers to take action. Customers are asked to follow the instruction in the Operator Manual for proper egress. Patients are not allowed to egress from the chair without assistance nor should they be left unattended. The chairs will be repaired on site; and, removal of the chair from service is not required with proper supervision and use. Customers are asked to complete the instructions for the acknowledgement form and return. The firm will contact the customers to arrange for repair. Questions regarding the field correction should be directed to the Customer Care department at 800-237-3377 or 352-854-2929 or e-mail QA-RA@wincomfg.com
- Initiated
- 2016-06-29
- Root cause
- Nonconforming Material/Component
- Terminated
- 2017-09-27
- Distribution
- Worldwide Distribution -- USA, including AR, CA, CO, CT, FL, GA, IL, KS, LA, MD, MI, MN, MO, MS, NC, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, VA, WA, WI, and Puerto Rico; and the countries of Canada, Australia, Italy, Panama, Switzerland and Taiwan.
- Recall number
- Z-0065-2017
- Recall status
- Terminated
- CDRH record ID
- 148797
- FDA record created
- 2016-10-10
- Recall event number
- 74922