Open original FDA record ↗Recall record
- FDA identifier
- Z-0065-2019
- Company in FDA record
- Vision RT Ltd
- FDA record creation date
- Oct 3, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
RTInterface error in the event of an I/O error between the IGC PCB and the Framegrabber PCB
Source facts
- Action
- Vision RT notified customers with Customer Information Bulletin dated August 24, 2018 to make them aware of the issue, how to spot it (in the unlikely event that it occurs) and how to deal with it. The customer contact also indicates that the issue is to be fixed with a firmware update and that this can be delivered remotely or installed on site by a service engineer,or transmit the firmware update to those customers whose systems can be accessed .
- Initiated
- 2018-08-24
- Root cause
- Software design
- Terminated
- 2022-07-08
- Distribution
- Nationwide Foreign: Australia, Austria, Azerbaijan, Belgium, Brunei, Bulgaria, Canada, China, Croatia, Denmark, Finland, Georgia, Germany, India, Israel, Italy, Japan, Korea, Lithuania, Malaysia, Morocco, Norway, Poland, Portugal, Puerto Rico, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Vietnam.
- Recall number
- Z-0065-2019
- Recall status
- Terminated
- CDRH record ID
- 167432
- FDA record created
- 2018-10-03
- Recall event number
- 81089