Skip to content

Public FDA record

Z-0065-2019: Integrated Gate Controller PCB (component of AlignRT Plus) Truebeam variant only. The AlignRT is a vide

Recall record Z-0065-2019 for Vision RT Ltd. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0065-2019
Company in FDA record
Vision RT Ltd
FDA record creation date
Oct 3, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

RTInterface error in the event of an I/O error between the IGC PCB and the Framegrabber PCB

Source facts

Action
Vision RT notified customers with Customer Information Bulletin dated August 24, 2018 to make them aware of the issue, how to spot it (in the unlikely event that it occurs) and how to deal with it. The customer contact also indicates that the issue is to be fixed with a firmware update and that this can be delivered remotely or installed on site by a service engineer,or transmit the firmware update to those customers whose systems can be accessed .
Initiated
2018-08-24
Root cause
Software design
Terminated
2022-07-08
Distribution
Nationwide Foreign: Australia, Austria, Azerbaijan, Belgium, Brunei, Bulgaria, Canada, China, Croatia, Denmark, Finland, Georgia, Germany, India, Israel, Italy, Japan, Korea, Lithuania, Malaysia, Morocco, Norway, Poland, Portugal, Puerto Rico, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Vietnam.
Recall number
Z-0065-2019
Recall status
Terminated
CDRH record ID
167432
FDA record created
2018-10-03
Recall event number
81089

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.