Z-0065-2020: Stryker Neurovascular, Merci 9F, 80cm, Balloon Guide Catheter, REF 90077, CE 0459, UPN Product No. 90077,

Recall record Z-0065-2020 for Stryker Neurovascular. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0065-2020
Company in FDA record
Stryker Neurovascular
FDA record creation date
Oct 8, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Labeling error; 8F Balloon Guide Catheters have been packaged within 9F Balloon Guide Catheter packaging

Source facts

Action
On 11/13/2018, a "Urgent Medical Device Voluntary Field Corrective Action" letter was sent to all affected consignees. Per the customer notification letter, it informed customers of the following: 1. Immediately check your internal inventory for 9F BGC Lot 0000010779. 2. Circulate this Customer Communication internally to all interested/affected parties. 3. If the specified lot is located, open the carton and remove the pouch. Without removing the device from the pouch or breaking the sterile barrier, inspect the device through the pouch and confirm that the text on the white strain relief reads 9F/.085 ID (shown in Fig. 1). If confirmed, no further action is needed to be taken for this device. 4. If the device identified above reads 8F/.078 ID (shown in Fig. 2) please discard the device and take note of this in the comments section in the attached response form. 5. On receipt of this form, replacement product will be issued to you. If you require assistance in executing these steps, please reach out to your Stryker sales representative. 6. Inform Stryker if any of the subject devices have been distributed to other organizations and provide Stryker the contact information for these organizations. 7. Please inform Stryker of any adverse events concerning the use of the subject devices. 8. Return the completed form to NVFieldActions@stryker.com. 9. We request that you respond to this notice within 7 calendar days from the date of receipt. The target date for completion of this action is 31 Jan 2019 and your timely response will enable us to ensure that we meet this target.
Initiated
2018-11-13
Root cause
Labeling False and Misleading
Terminated
2020-08-03
Distribution
OUS distribution only: CZECH REPUBLIC, ITALIA, Netherlands, Sweden, Germany, Austria, Israel, France, Belgium, Finland, Switzerland, Portugal, Finland
Recall number
Z-0065-2020
Recall status
Terminated
CDRH record ID
176254
FDA record created
2019-10-08
Recall event number
83735

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗