Skip to content

Public FDA record

Z-0066-2008: BD Microkeratome K3000 and K4000;Catalog Number: 378660, Becton Dickinson & Company (BD)

Recall record Z-0066-2008 for Becton Dickinson and Company. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0066-2008
Company in FDA record
Becton Dickinson and Company
FDA classification date
Oct 23, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Microkeratome blade may stop oscillating and transverse the eye causing injury to the cornea

Source facts

Action
BD Medical notified customers by Advisory Letter dated 8/09/07, advising users to return heads and bearing cartridges for replacement of the o-ring.
Initiated
2007-08-09
Root cause
Component design/selection
Terminated
2012-02-29
Distribution
Worldwide
Recall number
Z-0066-2008
Recall status
Terminated
CDRH record ID
54235
FDA classified
2007-10-23
Recall event number
38739

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗