Biological Indicator indicating a positive result does not maintain yellow color for the entire incubation period and some indicators experience media evaporation prior to the end of the labeled maximum incubation time.
Source facts
Action
An Important Product Labeling Correction - New Instructions for Use dated, June 26, 2009, has been proposed to be sent by the firm for Cyclosure Biological Indicators to consignees. The revised instructions reduce the maximum read time for the indicators to 3 days (72 hours). The letter also states why ASP changed the Instructions. Questions are to be directed to ASP Customer Care Center.
Initiated
2009-06-23
Root cause
Device Design
Terminated
2011-11-29
Distribution
Worldwide Distribution.
Recall number
Z-0066-2010
Recall status
Terminated
CDRH record ID
83495
FDA classified
2009-10-20
Recall event number
52632